SLAN®-96P Real-Time PCR System
FEATURES
Precise Temperature Control System:
1. The SLAN®system adopts Peltier technology which can accurately and quickly control increases and decreases of temperature with a precision of ±0.1℃ and features long service life, no noise and no pollution, etc.;
2. Multiple-point temperature control technology ensures the uniformity between different wells. The temperature correction technology guarantees the accuracy of temperature and thus ensures the accuracy of assay result;
3. PCR tube temperature-control technology guarantees the accuracy of actual temperature of reagent in the tube and enables the SLAN®to easily adjust for different reactions using different volumes.
Sensitive Photoelectric Detection System
1. High sensitivity photoelectric detector with no background noise which is fast and stable;
2. Lifetime maintenance-free ultra-bright LED cold light source with a large signal value and high stability;
3. Unique optical fiber conduction technology that greatly improves fluorescence collection efficiency and avoids fluorescence interference between neighboring tubes;
4. The system automatically selects best gain value for each tube without the need for user configuration. The SLAN®has an extremely broad measuring range for fluorescence;
5. Extremely high detection sensitivity with an extremely low fluorescence background value. It’s not necessary for users to make optical correction or background correction.
SLAN-96P Multi-tasking Software Interface
(analysis screen)
Unique Electronic Automatic Hot Lid System
1. Flexible heating technology ensures uniform hot lid temperature;
2. The electronic automatic hot lid has an adjustment function for different 0.2ml reaction tubes to ensure even tube pressure;
3. Once the assay starts, the hot lid has automatic locking function to prevent assay failure following accidental opening.
Other functions
· Power-off protection | · Automatic saving result | · Export of result |
· Real-time data display | · Quantitative and qualitative analysis | · File encryption |
· Automatic baseline optimization | · Digital filtering | · Search function |
· Report sheet print and edit | · Melt curve function |
SPECIFICATIONS
Product model | SLAN-96P |
Sample capacity | 96 wells (dual 48/48 reaction module) |
Sample size | 15-50mL |
Consumable | 0.2mL PCR tubes, 8-tube strips, 48-well plates |
Parallel running mode | Dual reaction blocks/running 2 tests independently |
Temperature range | 4-99°C |
Max ramp rate | 4.0°C/sec |
Temperature accuracy | ±0.1°C |
Temperature uniformity | ±0.1°C |
Thermal cycling system | Peltier-based system |
Temperature control mode | Tube control/block control |
Light Source | LED (maintenance-free) |
Detector | High sensitivity photoelectric sensor |
Sensitivity | 1 copy |
Linearity range | 100-1010 |
Resolution | Can discriminate between 1000 copies and 2000 copies |
Excitation | CH1 470nm CH2 530nm CH3 580nm
CH4 630nm CH5/CH6 custom-made |
Emission | CH1 510nm CH2 565nm CH3 620nm
CH4 665nm CH5/CH6 custom-made |
Dyes and probes | CH1 FAMTM /SYBRGreen® CH2 VIC®/HEX/J0ETM /TET
CH3 ROX/TexasRed® CH4 CY5TM |
Hot-lid | 30°C~108°C (default105°C, adjustable) electronic automatic hot-lid |
Power supply | 230VAC, 50Hz |
Power consumption | 850VA |
Dimensions | 380mm x 520mm x 250mm (WxDxH) |
Weight | 18Kg |
Communication | RS232, USB |